Made In The U.S.A. Since 1938

Lumenite Control Technology, Inc.

Engineers • Designers • Manufacturers

• Stainless steel welded tube and socket

• 4” standard case size

• ± 1% full scale calibrated accuracy

• 100% helium leak tested

• Electropolished 316 s.s. diaphragm

• Available in 1.5 or 2” Tri-Clamp Connection

• 100% Helium leak tested

• 304 s.s. housing

• Ambient temperature range up to 300˚F standard•

• 100% hand calibrated and certified to exceed the current ASME B40.1

• 90mm standard case

• Calibrated Accuracy = 1.5% f.s. for ranges 100 p.s.i. or more

• All gauges designed for C.I.P.

• Dry gauges suitable for autoclave sterilization

• Designed to withstand 50% overpressure with minimal loss of accuracy

Pressure Instrumentation

        Lumenite® sanitary analog pressure gauges,  digital pressure gauges / transmitters and sanitary pressure transmitters conform to 3A standard 37-01.  All Lumenite® sanitary products are TPV (third party

verified).  SCROLL DOWN TO VIEW !!!

Model LSPG  Sanitary Analog Pressure Gauge

Available Connections : 1.5 & 2" Tri-Clamp Connections

Model LSDPG  Sanitary Digital Pressure Gauge / Transmitter

Available Connections : 1.5 & 2" Tri-Clamp Connections

• 3A, CEE, TPV third party verified

• TPA terminal point accuracy (no BFSL)

• 5 digit LCD display with .60 inch characters and optional back lighting

• 0.25% full scale accuracy

• Optional 4-20mA transmitter output with battery backup

• Replaces both a pressure gauges and 4-20mA pressure transmitter

• 316 s.s. sanitary fitting and socket for CIP cleaning

• Percent of range bargraph

• Vacuum through 20,000 psi. ranges

• Alternate metric and English unit of measure,  programmable through

  keypad,  include mmHg, in H20, mBar, Kpa, kg/cm2, and bar

• Password protected zero, midpoint, and span

• Lower 6’o’clock standard mount

• Operating temperature 14˚F to 140˚F (-10˚C to 60˚C)

Model LSPT  Sanitary Pressure Transmitter

Available Connections : 1.5 & 2" Tri-Clamp Connections

• Designed for food and pharmaceutical applications

• Meets all 3A, USDA and TPV third party verification testing / evaluation criteria

• Accuracy : 0.50% BFSL

• Linearity : ±0.08 BFSL or better

• Hysteresis : 0.15% or better

• Repeatability : 0.07%

• Stability : ±0.2% of calibrated rage for 1 year

• Over-Range Rating : 2 times base range at burst of 800%

• Zero/Span Adjustments : ±10%

• Process Temperature Range : -20˚F to 300˚F

• Storage Temperature : -65˚F to 250˚F

• Operating Temperature : -0˚F to 180˚F

• Durability : 10 to the 7th cycles with negligible performance change

• Humidity Effect : no performance effect at 95% relative humidity

  non-condensing

• Vibration Sweep : less than ±0.1% F.S. effect for0-2000HZ at 20 g’s

  in any axis

• Shock : less than ±0.05% F.S. effect for 100g’s 20ms shock in any axis

• Position Effect : less than 0.1% F.S.

• Output : 4 to 20mA standard output; optional 1-5VDC, 1-6VDC reverse

  polarity protected

• Housing : highly polished 304 s.s.; product contact surfaces 316 s.s.

• Approvals : meets ALL 3A, USDA and FDA requirements